• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH VIPER PRIME CFX X-TAB POLYAXIAL SCREW 5.5 7 X 35MM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH VIPER PRIME CFX X-TAB POLYAXIAL SCREW 5.5 7 X 35MM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Model Number 186770335S
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Code Nonunion/Delayed-union Bone Fracture (2369)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d2b: additional device product codes: kwq, nkb.D9: complainant part is not expected to be returned for manufacturer review/investigation.E1: initial reporter facility name: (b)(6) hospital.E3: reporter is a j&j employee.H3, h4, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot: a manufacturing record evaluation was performed for the finished device.Product code: 186770335s; lot number: tbahfi; it was electronically reviewed and no nonconformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 13/12/2021; manufacturing site:jabil le locle; expiry date:31/10/2026.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in japan as follows: it was reported that on (b)(6) 2022, a tlif (l5/s1) was performed for spinal stenosis.A revision surgery was performed in (b)(6) 2023 due to loosening of the screws and non-union between l5/s1.In the revision surgery, the cage was removed and replaced with a non-synthes product.The screws were removed and replaced with screws of a larger size.The surgery was completed successfully, and the patient outcome was reported to be stable.This report involves one viper prime cfx x-tab polyaxial screw 5.5 7 x 35mm.This is report 4 of 4 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER PRIME CFX X-TAB POLYAXIAL SCREW 5.5 7 X 35MM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
SZ  
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 
SZ  
3035526892
MDR Report Key17266237
Report Number1526439-2023-01231
Device Sequence Number12239895
Product Code KWP
UDI-Device Identifier10705034530161
UDI-Public(01)10705034530161
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K173095
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number186770335S
Device Catalogue Number186770335S
Device Lot NumberTBAHFI
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/13/2023
Initial Report FDA Received Date07/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
T-PAL PROTI, 10X9X28 MM.; VIPER PRIME CFX XTAB 7X35MM TI.; VIPER PRIME CFX XTAB 7X35MM TI.; VIPER PRIME CFX XTAB 7X35MM TI.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-