MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION DYNAMIC XT; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION DYNAMIC XT; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number 86706
Medical Device Problem Code Material Perforation (2205)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/12/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during preparation for a procedure using a dynamic xt catheter the shaft of the sheath was broken.They exchanged the catheter, and the issue was resolved.The procedure was then completed with no patient complications.The catheter is expected to be returned for analysis.
 
Additional Manufacturer Narrative
Upon receipt at boston scientific's post market laboratory the catheter was visually inspected, and it was found the plunger that controls the steering of the catheter had separated from the rest of the handle.The reported allegation of irrigation failure was confirmed, however, the root cause for the break could not be determined from the information provided from the site or from examining the damaged device.
 
Event or Problem Description
It was reported that during preparation for a procedure using a dynamic xt catheter the shaft of the sheath was broken.They exchanged the catheter, and the issue was resolved.The procedure was then completed with no patient complications.The catheter has been received for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAMIC XT
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key17270174
Report Number2124215-2023-34168
Device Sequence Number7911776
Product Code DRF
UDI-Device Identifier08714729877622
UDI-Public08714729877622
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K921872
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/04/2025
Device Model Number86706
Device Catalogue Number86706
Device Lot Number0030486967
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/09/2023
Supplement Date Received by Manufacturer08/24/2023
Initial Report FDA Received Date07/06/2023
Supplement Report FDA Received Date09/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-