• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US HEALIX KNOTLESS AWL, 3.4MM; BONE AWL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US HEALIX KNOTLESS AWL, 3.4MM; BONE AWL Back to Search Results
Model Number 219495
Medical Device Problem Codes Degraded (1153); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: udi: (b)(4).D4: the expiration date is currently unavailable.E3: reporter is a j&j sales representative.H4: the device manufacture date is currently unavailable.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported by the sales rep that during a rotator cuff repair procedure on (b)(6) 2023, it was observed that the tip on the healix knotless awl, 3.4mm device was blunt and looked bent.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary the complaint device was received and evaluated.Upon visual inspection the device showed wear use.It could be observed the tip of the device with a bent condition.Also, it could be identified the etch marks slightly faded.A black mark near the tip was identified.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.Based on the visual inspection results, this complaint can be confirmed.This condition has possibly occurred after using it for many procedures.The possible root cause for the condition of the device could be related to heavily use; moreover, drop of the unit, mishandling or procedural variables that could cause the bent condition of the device.Since this device is reusable, the continuous use and sterilization process can cause worn condition.However, it cannot be conclusively affirmed.As per ifu, inspect for damage prior to use.Replace a damaged, worn or bent instrument.Do not attempt to straighten, sharpen or repair.Do not use a depuy mitek reusable instrument for any purpose other than its intended use as that may result in damage to the instrument.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEALIX KNOTLESS AWL, 3.4MM
Common Device Name
BONE AWL
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key17271074
Report Number1221934-2023-02635
Device Sequence Number3567336
Product Code LXH
UDI-Device Identifier10886705022373
UDI-Public10886705022373
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number219495
Device Catalogue Number219495
Device Lot Number1211001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/25/2023
Supplement Date Received by Manufacturer07/21/2023
08/09/2023
Initial Report FDA Received Date07/06/2023
Supplement Report FDA Received Date07/25/2023
08/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-