| Catalog Number |
07251670190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The sample was requested for investigation.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they received questionable results for samples from one patient tested with the elecsys anti-ccp assay on a cobas e 801 analytical unit (serial number (b)(6)) and compared to the siemens method.An interference is suspected by the customer.On (b)(6)2023, a sample from the patient was tested using the elecsys anti-ccp assay, resulting in a value of 309 u/ml.On (b)(6) 2023, a sample from the patient was tested using the elecsys anti-ccp assay, resulting in a value of 249 u/ml.On (b)(6) 2023, a sample from the patient was tested using the siemens anti-ccp method, resulting in a value of <0.5 ku/l.This result is consistent with the patient¿s clinical picture.
|
| |
|
Additional Manufacturer Narrative
|
|
A sample of the patient was provided for investigation.The high elecsys anti-ccp result of the customer could be reproduced.No interfering factor against the streptavidin component of the assay could be determined in the sample.Based on the investigation results, the elecsys anti-ccp result is considered to be true positive.The investigation did not identify a product problem.
|
| |
|