• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. NV GXL LINER LIPPED 40MM ID, GROUP 3 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. NV GXL LINER LIPPED 40MM ID, GROUP 3 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number NV GXL LINER LIPPED 40MM ID, GROUP 3 CUPS
Medical Device Problem Codes Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Osteolysis (2377)
Date of Event 06/13/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Section d10: concomitant products biolox delta femoral head 40mm 0d, +3.5mm (cat# 170-40-03 / serial# (b)(6)).Alteon 6.5mm screw, 35mm (cat# 180-65-35 / serial# (b)(6)).Integrip cc, cluster 58mm, g3 (cat# 186-01-58 / serial# (b)(6)).Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
As reported, approximately seven years post initial left tha, the 71 y/o male patient presented with complaints of pain and dissatisfaction with their primary exactech hip.The patient had a recall gxl poly liner and was booked for revision surgery.The patient underwent total acetabular revision to a exactech 28mm cocr +10mm head and competitor's devices.There was no breakage of device.There was a 30-45 mins surgical delay/prolongation.Patient was last known to be in stable condition following the event.X-ray and images received.The devices are not available for evaluation due to the products were sent to the lab at the hospital.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: h6.Mdr section codes updated/corrected: a, b, c, d, e, g.The most likely underlying cause for the revision reported is prosthesis wear and a combination of risk factors: such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential).Additional reasons for the revision may be acetabular loosening.However, this cannot be confirmed from the reported information and the devices were not available for evaluation.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NV GXL LINER LIPPED 40MM ID, GROUP 3 CUPS
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key17274412
Report Number1038671-2023-01572
Device Sequence Number5086188
Product Code JDI
UDI-Device Identifier10885862207043
UDI-Public10885862207043
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K121392
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 07/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/13/2020
Device Model NumberNV GXL LINER LIPPED 40MM ID, GROUP 3 CUPS
Device Catalogue Number132-40-53
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/14/2023
Supplement Date Received by Manufacturer07/08/2025
Initial Report FDA Received Date07/06/2023
Supplement Report FDA Received Date07/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/16/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1732-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
3.2MM DRILL BIT30MM 1PK.; ALTEON 6.5MM SCREW, 35MM.; WEDGE PLASMA X/O SZ 13.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexMale
-
-