| Model Number |
V272 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Failure to Capture (1081); Over-Sensing (1438); Pacing Problem (1439); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/27/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, that the patient with this cardiac resynchronization therapy pacemaker (crt-p) exhibited loss of capture (loc) or pacing inhibition.Upon review, the crt-p was found in safety mode.Subsequently, this device was explanted and replaced.No additional adverse patient events were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode due to system resets.It was confirmed that brady therapy remained available.The system resets other transient elevated power states and caused the device to enter safety mode.Engineers determined that this device was demonstrating behavior consistent with a high internal cell impedance within the battery.The high internal impedance resulted in the system resets due to temporary high-power consumption and subsequent reversion to safety mode operation.Boston scientific has issued a field safety notice regarding ingenio extended life (el) and cardiac resynchronization therapy pacemaker (crt-p) devices that may exhibit this behavior.This particular device is included in the high impedance in ingenio extended life (el) pacemaker and crt-p advisory population.
|
| |
|
Event or Problem Description
|
|
It was reported that, that the patient with this cardiac resynchronization therapy pacemaker (crt-p) exhibited loss of capture (loc) or pacing inhibition.Upon review, the crt-p was found in safety mode.Subsequently, this device was explanted and replaced.No additional adverse patient events were reported.
|
| |
|