EXACTECH, INC. NV GXL LNR, +5LAT, 36MM G2-52/54MM CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER
|
Back to Search Results |
|
| Model Number |
136-36-52 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Osteolysis (2377)
|
| Date of Event |
04/17/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Pending investigation.D10.Concomitants: 4343627 170-36-00 - biolox delta femoral head 36mm od, +0mm.4284798 180-01-52 - nv crown cup clstr hole 52mm group 2.4317553 180-65-20 - alteon 6.5mm screw, 20mm.3706124 180-65-35 - alteon 6.5mm screw, 35mm.4291326 188-00-08 - wedge plasma s/o sz 8.
|
| |
|
Event or Problem Description
|
|
As reported via legal documentation, a patient had left hip replacement surgery on (b)(6) 2016.They subsequently underwent left hip revision surgery on (b)(6) 2023, approximately 6 years 10 months after their primary procedure.(b)(6) 2023 op report states that there was some mild osteolysis around the rim of the acetabulum, but the stem was well fixed and retained.The trunnion of the femoral head was noted to be in good condition.The acetabular liner had significant wear, some cracking along the superior rim of the acetabulum.The cup was well fixed and in good position.The central hole and around the rim did not indicate significant osteolytic defects; bone grafting was not needed.There is no other patient demographic or medical history available.There is no device return.There are no photos or other images of the device provided.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H10.Updated/additional information ¿ d5.G1.G2.G4.H6.The most likely cause for the revision reported due to prosthesis wear is a combination of risk factors including, use error, implant positioning, implant size selection, and patient factors may have also been a contributing factor to the early prosthesis wear.However, this cannot be not confirmed with the information provided; devices were not returned.There is no other information available.These devices are used for treatment not diagnosis.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, e4 (remove erroneous entry from initial report), h6: clinical code, component code, type of investigation and investigation findings.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|