Model Number 22295-115
Medical Device Problem Codes Leak/Splash (1354); Loose or Intermittent Connection (1371); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3- date of event: estimated the device was not returned for analysis.A review of the lot history record and similar incident review of the reported lot could not be conducted because the lot number was not provided.It is possible that the guide wires were not coaxially aligned during insertion thus resulting in the reported difficult to advance; however this cannot be confirmed.Additionally, it is possible the devices were not fully tightened/closed resulting in the reported loose or intermittent connection and the reported leak; however as the device was not returned for analysis this cannot be confirmed.The investigation was unable to determine a conclusive cause for the reported difficulties.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported in a general comment that the guide wire accessory kit with 115 rhv (rotating hemostatic valve) was noted to not close properly, not achieving complete hemostasis and blood comes out.Also when the valve is completely open, it is very difficult to pass two 0.014 guide wires through since there's resistance when advancing two at once.However, the the procedure was finished with the same rhv.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUIDE WIRE ACCESSORY KIT
Common Device Name
ACCESSORIES
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key17278144
Report Number2024168-2023-07185
Device Sequence Number17423683
Product Code DQX
UDI-Device Identifier08717648079627
UDI-Public08717648079627
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K950752
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number22295-115
Device Catalogue Number22295-115
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/29/2023
Initial Report FDA Received Date07/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
0.014 GUIDE WIRE
Patient SexUnknown
-
-