| Model Number |
0276 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Low impedance (2285); Defective Device (2588); Material Integrity Problem (2978)
|
| Health Effect - Clinical Codes |
Electric Shock (2554); Shock from Patient Lead(s) (3162)
|
| Date of Event |
12/06/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) and right ventricular (rv) defibrillation lead exhibited significant noise with oversensing since december 2016, resulting in an inappropriate shock and inappropriate atp.More recently, this oversensed noise resulted in an inappropriate burst of atp and greater than two seconds of pacing inhibition.It was unclear whether the patient was pacer dependent as the device rv paced 1%.Technical services (ts) reviewed possible causes and troubleshooting options, and possible lead integrity concerns were discussed.This icd system remains in service, however further evaluation was recommended.No additional adverse patient effects were reported.
|
| |
|