Model Number 72201541
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Perforation (2001)
Date of Event 06/19/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H10: internal complaint reference: case (b)(4).
 
Event or Problem Description
It was reported that during an arthroscopy, the bioraptor suture anchor broke while tapping into the bone.The broken pieces were successfully removed from the patient with the help of grasper.The procedure was completed with non-significant surgical delay using a back-up device in an additionally drilled bone hole.No further complications were reported.
 
Additional Manufacturer Narrative
H10: h3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found that tensile strength requirements are specified.A material certificate of analysis (coa) is required for raw material.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIRPTR 2.3 PK SUTURE ANCHOR W/ ULTRAB
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key17278958
Report Number1219602-2023-01084
Device Sequence Number17463139
Product Code MBI
UDI-Device Identifier03596010595072
UDI-Public03596010595072
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K071586
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/04/2026
Device Model Number72201541
Device Catalogue Number72201541
Device Lot Number2080559
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/19/2023
Supplement Date Received by Manufacturer09/07/2023
Initial Report FDA Received Date07/07/2023
Supplement Report FDA Received Date09/11/2023
Date Device Manufactured10/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age36 YR
Patient SexFemale
-
-