| Model Number |
20472100 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturer's ref# (b)(4).This is an initial report.Investigation is still in progress.A final report will be submitted upo completion.
|
| |
|
Event or Problem Description
|
|
Estimated date of incident: (b)(6) 2023.It was reported that the charger port melted.User says it had gotten a little warm while charging lately.No harm to people or property.Original equipment was used.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's ref# (b)(4).This is an initial report.Investigation is still in progress.A final report will be submitted upo completion.28jul2023.Technical investigation concluded: a device history record review have been completed.No deviation or changes were found during manufacturing of the device.Trended case concluded: the connector is broken, probably because the user has (not intentionally) misused it, while inserting the cable.If the cable has been slightly out of specification that could have had a negative effect on this.This has caused a low ohmic connection in the cable connector - almost a short circuit - that will emit heat when the cable is inserted into the wall charger, causing up to the rated 5 w of the wall charger to be dissipated inside the connector.Micro usb is a very well known and used standard but can mechanically break down.The clinical investigation concluded: it was reported that the charger melted at the port while the charger was plugged in, charging the hearing aids.User had previously noticed that it had been getting a little warm while charging.There was no harm to the user.There is no mention of property damage.Original accessories were used.The chargers have been tested according to applicable safety standards.These risks are reduced as far as possible and the probability of occurrence of these events is considered very low.In case of moisture, sweat or dirt in the connector, there can be high temperatures inside the connector due to power dissipation when connected to the charger plug.There can be melting of the plastic around the connector and is visible to the naked eye.It is highly unlikely that the user would touch in case there appears to be melting or high temperatures.In the unlikely event of the user touching this overheated device, it can lead to superficial or minor injury that would in most cases require no medical intervention.There are no new clinical aspects (e.G., unforeseen use or clinical conditions) discovered from this event.Therefore, it is concluded that the current clinical evaluation and risk/benefit conclusion herein are sufficient.Risk register conclusion reference: the residual risk after the risk control measure(s) is acceptable.No new hazardous situations revealed and no need to update the risk register.Manufacturer's investigation is final.This is a final report.
|
| |
|
Event or Problem Description
|
|
Estimated date of incident (b)(6) 2023.It was reported that the charger port melted.User says it had gotten a little warm while charging lately.No harm to people or property.Original equipment was used.28jul2023.No further follow up received.
|
| |
|