• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC 14 CM QD ANGLE ATTACHMENT; MOTOR, DRILL, ELECTRIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC 14 CM QD ANGLE ATTACHMENT; MOTOR, DRILL, ELECTRIC Back to Search Results
Model Number QD14-G1
Medical Device Problem Code Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/06/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the attachment device was heating up.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available for use.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi:(b)(4).
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition of the device heating up was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation, it was determined that the device had vibration.Upon disassembling, the locking components were identified to be damaged.It was further determined that the device failed pretest for vibration assessment.The assignable root cause was determined to be traced to maintenance, which is improper maintenance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
14 CM QD ANGLE ATTACHMENT
Common Device Name
MOTOR, DRILL, ELECTRIC
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 33410
3035526892
MDR Report Key17279040
Report Number1045834-2023-01037
Device Sequence Number3566662
Product Code HBC
UDI-Device Identifier00845384016458
UDI-Public00845384016458
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K133604
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberQD14-G1
Device Catalogue NumberQD14-G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2023
Supplement Date Received by Manufacturer09/22/2023
Initial Report FDA Received Date07/07/2023
Supplement Report FDA Received Date09/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-