| Model Number |
AF5040MB |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
Low Oxygen Saturation (2477)
|
| Date of Event |
02/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Face masks inside the kit are flat (underinflated).
|
| |
|
Additional Manufacturer Narrative
|
|
The mask deflating and not properly sealing to patient would cause a disruption in resuscitation efforts and put them at risk.
|
| |
|
Event or Problem Description
|
|
Face masks inside the kit are flat(underinflated).
|
| |
|
Additional Manufacturer Narrative
|
|
The mask deflating and not properly sealing to patient would cause a disruption in resuscitation efforts and put them at risk.Summary: complaint confirmed via returned devices.The masks are able to be inflated and used properly.The customer was notified of potential causes for deflated masks.Ra: instructions for mask use in combination with rma-20020 are sufficient risk analysis for the reported product complaint.
|
| |
|