This report is related to mdr number 3011632150-2023-00086.There was no reported device malfunction and the device was not returned for analysis.A review has been performed of the manufacturing certificate of conformance associated with this lot.There were no non-conformances or deviations which could be associated with this event.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.The reported patient effects of restenosis and claudication/leg pain are listed in the biomimics 3d instructions for use and are known patient effect of peripheral stenting procedures.There is no indication of a product quality issue with respect to the design, manufacture or labelling of the device.If further information regarding this event becomes available, a follow-up report will be submitted.
|
|
This report is related to mdr number 3011632150-2023-00086.The patient was treated as part of the mimics 3d usa post-market observational study.On the (b)(6) 2021, the patient was implanted with two biomimics 3d (bm3d) stents, a 7.0 x 150 mm stent and a 7.0 x 80 mm stent (the subject of this report) which were used to treat a denovo lesion of the superficial femoral artery (sfa) middle third to proximal popliteal artery in the right leg.A retrograde approach was used and the lesion was prepared using pre-dilation with percutaneous transluminal angioplasty (pta) balloon and atherectomy.The treated segment was post-dilated with pta.The site identified a restenosis of treated segment (target lesion) on (b)(6) 2022.It was reported as definitely related to the device and not related to the procedure.It was also reported as target lesion related.The patient had a 12-month ankle brachial index (abi) and duplex ultrasound (dus) on the (b)(6) 2022.The abi on the right was measured at 0.37 (resting) and dus showed monophasic waveforms throughout and elevated velocities.The patient was experiencing symptoms consistent with rutherford class 3.A diagnostic angiogram on (b)(6) 2022 identified that the right mid/distal sfa had a 99% target lesion (tl) in-stent restenosis (isr).The percutaneous intervention was conducted on (b)(6) 2022.The patient underwent laser atherectomy and drug coated balloon / drug eluting balloon (dcb/deb) treatment on the sfa distal third.This left less than 10% residual stenosis.The intervention was reported as a target lesion/vessel revascularisation (tlr/tvr).The outcome was reported as resolved/recovered.The devices remain implanted.
|