TRIVIDIA HEALTH INC SYRINGES; PISTON SYRINGE Back to Search Results
Model Number SYR, GNP .3CC 31G 100CT5/16" 8/CS
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2023
Type of Reportable Event Malfunction
Event or Problem Description
Consumer reported complaint for the trueplus single-use insulin syringes.Customer stated that the 31g syringes did not aspirate the insulin ( levemir).Customer stated this just started within the last couple of weeks.The package had not been open or damaged when received by the customer.The customer did not claim they were injured using the syringes.The customer feels well and did not report any symptoms.No medical attention associated with the use of the product was reported.
 
Additional Manufacturer Narrative
Internal report reference number: (b)(4).Syringes were not returned for evaluation.Note: manufacturer contacted customer in a follow-up call on (b)(6) 2023 to ensure the replacement products resolved the initial concern - able to establish contact with customer who stated replacement products resolved initial concern.Customer stated he had discarded the alleged defective syringes.
 
Additional Manufacturer Narrative
Sections with additional information as of 08-aug-2023: h6: updated fda's type of investigation, investigation findings, and investigation conclusions.H10: complaint was forwarded to supplier quality based on complaint's description for investigations.No product was returned to thi.Internal evaluation has been completed by the manufacturer.No abnormalities observed with retention samples.Most likely underlying root cause: mlc-009: use error caused or contributed to event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGES
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
TRIVIDIA HEALTH INC
2400 nw 55th court
fort lauderdale FL 33309
Manufacturer Contact
karen devincent
2400 nw 55th court
fort lauderdale, FL 33309
MDR Report Key17280386
Report Number1000113657-2023-00366
Device Sequence Number3566714
Product Code FMF
UDI-Device Identifier00021292005863
UDI-Public(01)00021292005863
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K993017
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 08/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/19/2025
Device Model NumberSYR, GNP .3CC 31G 100CT5/16" 8/CS
Device Lot NumberNP22192
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Distributor Facility Aware Date06/13/2023
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/13/2023
Supplement Date Received by Manufacturer07/24/2023
Initial Report FDA Received Date07/07/2023
Supplement Report FDA Received Date08/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-