• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OCTRODE LEAD KIT, 30CM LENGTH; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OCTRODE LEAD KIT, 30CM LENGTH; SCS LEAD Back to Search Results
Model Number 3183
Medical Device Problem Code Low impedance (2285)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 03/01/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event is estimated.
 
Event or Problem Description
Related manufacturer reference number: 1627487-2023-03181.Related manufacturer reference number: 1627487-2023-03182.Related manufacturer reference number: 1627487-2023-03183.It was reported that the patient never experienced effective therapy from their occipital scs system for their migraines.The ipg for their occipital system in inoperable.Additionally, the patient experienced ineffective therapy from their lumbar scs system due to high and low impedances.Reprogramming was unable to resolve the issue.As such, surgical intervention may take place at a later date to address the issue.Investigation was unable to determine which of the occipital leads attributed to the ineffective therapy.
 
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event or Problem Description
Additional information received indicates that surgical intervention took place on (b)(6) 2023 wherein the occipital system was explanted to address the issue with this system.Additionally, the lumbar system was also explanted and replaced with a new system to address the issue with the lumbar system.Reportedly, therapy was restored post operatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 30CM LENGTH
Common Device Name
SCS LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
heidi syndergaard
6901 preston road
plano, TX 75024
9723098000
MDR Report Key17280948
Report Number1627487-2023-03184
Device Sequence Number19703098
Product Code LGW
UDI-Device Identifier05414734401692
UDI-Public05414734401692
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/28/2014
Device Model Number3183
Device Catalogue Number3183
Device Lot Number3870884
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/12/2023
Supplement Date Received by Manufacturer09/15/2023
Initial Report FDA Received Date07/07/2023
Supplement Report FDA Received Date10/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/27/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCS ANCHOR (X2); SCS IPG
Outcome Attributed to Adverse Event Other;
Patient Age52 YR
Patient SexMale
Patient Weight73 KG
-
-