ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number AGFN-756
Medical Device Problem Codes Material Frayed (1262); Device Damaged Prior to Receipt by User (1570)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/18/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2023, a 21mm sjm regent mechanical heart valve (flexcuff), was selected for an implant.The surgeon removed the valve from the operating room warehouse, opened the package and found that the valve ring was partially damaged.The physician did not want to waste the valve and thought that implantation would not affect the performance of the valve.The physician attempted implanting it in the patient but removed it due to difficulty suturing.A new non-abbott valve was used to complete the procedure.The product box containing the device was also noted to have been damaged.The patient is stable.
 
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Additional Manufacturer Narrative
The reported event of damaged packaging and device damage was reported.The investigation confirmed that the sewing cuff was torn/frayed and contained blood.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Image from the field appeared to show the sweing cuff.Based on the information received and returned device analysis, the reported damaged packaging and device damage could not be conclusively determined.However, incident is consistent with damage during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REGENT HEART VALVE WITH FLEX CUFF
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key17282830
Report Number2135147-2023-02939
Device Sequence Number19703116
Product Code LWQ
UDI-Device Identifier05414734005975
UDI-Public05414734005975
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/04/2025
Device Model NumberAGFN-756
Device Catalogue Number21AGFN-756
Device Lot Number7613715
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2023
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/15/2023
Supplement Date Received by Manufacturer07/07/2023
Initial Report FDA Received Date07/07/2023
Supplement Report FDA Received Date07/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/05/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age52 YR
Patient SexMale
Patient Weight56 KG
-
-