Model Number VERSION 1.0
Medical Device Problem Codes Retraction Problem (1536); Material Deformation (2976); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/15/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: as the lot number for the device was not provided, a review of the device history records could not be performed.The return of the sample is pending.However, photos, images and videos were provided for review.The investigation of the reported event is currently underway.H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a recanalization procedure, the tip of catheter was allegedly out of shape after first round of use.It was further reported that device cannot remove from the sheath.The procedure was completed by using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing-related cause for this event.Investigation summary: the sample was returned for evaluation.A catheter was physically investigated.The catheter was sent with guidewire and introducer sheath connected and stuck together.The tube kink at 22 cm distance from the tip of the catheter was noted.The tip of the tube was melted on the helix.The coiled tip of the guidewire was missing, and guidewire was delaminated at the place of the tube kink (22 cm).The inside of the catheter and helix were heavily blocked with the coagulated material.It was not possible to run the catheter due to blockage.Therefor, the investigation is confirmed for the reported issues.A clear root cause could not be identified but a blockage of the catheter represents a known inherent risk.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a recanalization procedure, the tip of catheter was allegedly out of shape after first round of use.It was further reported that device cannot remove from the sheath.The procedure was completed by using another device.There was no reported patient injury.
 
Event or Problem Description
It was reported that during a recanalization procedure, the tip of catheter was allegedly out of shape after first round of use.It was further reported that device cannot remove from the sheath.The procedure was completed by using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: as the lot number for the device was provided, a review of the device history records will be performed.The return of the sample is pending.However, photos, images and videos were provided for review.The investigation of the reported event is currently underway.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAREX
Common Device Name
THROMBECTOMY & ATHERECTOMY
Manufacturer (Section D)
UNKNOWN
BR 
Manufacturer (Section G)
UNKNOWN
BR  
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key17285606
Report Number3008439199-2023-00116
Device Sequence Number3023394
Product Code MCW
UDI-Device Identifier07640142810292
UDI-Public(01)07640142810292
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K211738
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVERSION 1.0
Device Catalogue Number80201
Device Lot Number221844
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/15/2023
Supplement Date Received by Manufacturer07/07/2023
09/14/2023
Initial Report FDA Received Date07/10/2023
Supplement Report FDA Received Date07/30/2023
09/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-