• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON INSERT CS #6 10MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON INSERT CS #6 10MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5531-G-610-E
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/12/2023
Type of Reportable Event Malfunction
Event or Problem Description
Primary procedure, left mako knee.It was reported that after impaction, one side of the locking wire was observed to be above the rim of the tibial baseplate.The wire appeared to have been molded in the opposite direction it should have been.Mps confirmed that no excessive force was used during impaction (impaction consisted of 1 impact), and that impaction was not performed anywhere close to the wire.Insert was removed and a second implant was used to complete the surgery successfully with a surgical delay of less than a minute.Mps has provided intra-op pictures of the implant, and confirmed that no further information will be released by the hospital or surgeon.
 
Additional Manufacturer Narrative
Reported event: an event regarding disassociation involving a triathlon insert was reported.The event was confirmed.Method & results: product evaluation and results: visual inspection: the reported device was not returned however photographs were provided for review.The photographs show an implanted insert with a disassociated locking wire.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that after impaction, one side of the locking wire was observed to be above the rim of the tibial baseplate.The reported device was not returned however photographs were provided for review.Visual inspection: the photographs show an implanted insert with a disassociated locking wire.No further investigation for this event is possible at this time as no devices was received by stryker orthopaedics.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X3 TRIATHLON INSERT CS #6 10MM
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key17286337
Report Number0002249697-2023-00757
Device Sequence Number19588703
Product Code MBH
UDI-Device Identifier07613327390575
UDI-Public07613327390575
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K173849
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5531-G-610-E
Device Catalogue Number5531-G-610-E
Device Lot Number1J2TN6
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/12/2023
Initial Report FDA Received Date07/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/02/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-