• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 14 DAY; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 14 DAY; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71938-01
Medical Device Problem Code Product Quality Problem (1506)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/19/2023
Type of Reportable Event Malfunction
Event or Problem Description
Customer reports their adc device is too hot to hold.No further details are available at this time.It is unknown if the customer's device was too hot to hold during use or while charging.There was no report of fire, smoke, explosion, or shock.It is unknown at this time if the charging cable and adapter used was the cable and adapter supplied by adc for use with the libre device.There was no report of death, serious injury, or mistreatment associated with this event.Issues involving thermal events occurring with adc freestyle libre and libre 2 readers is associated with report of corrections and removal number 2954323-02/09/23-001-c, submitted to the fda on 09-feb-23.Adc field action fa1005-2023 was issued to the us 09feb23.
 
Additional Manufacturer Narrative
The customer¿s product has been requested for investigation.A follow-up report will be filed once additional information is obtained.All pertinent information available to abbott diabetes care has been submitted.
 
Additional Manufacturer Narrative
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The dhrs (device history review) for the libre reader were reviewed and the dhrs showed the libre reader passed all tests prior to release.In addition, the review determined that the correct cable and adapter were part of the reader kit pack and there was no indication that the product did not meet specifications.If the product is returned, the case will be re-opened, and a physical investigation will be performed.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
Customer reports their adc device is too hot to hold.No further details are available at this time.It is unknown if the customer's device was too hot to hold during use or while charging.There was no report of fire, smoke, explosion, or shock.It is unknown at this time if the charging cable and adapter used was the cable and adapter supplied by adc for use with the libre device.There was no report of death, serious injury, or mistreatment associated with this event.Issues involving thermal events occurring with adc freestyle libre and libre 2 readers is associated with report of corrections and removal number 2954323-02/09/23-001-c, submitted to the fda on 09-feb-23.Adc field action fa1005-2023 was issued to the us 09feb23.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 14 DAY
Common Device Name
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key17286692
Report Number2954323-2023-29479
Device Sequence Number6214276
Product Code PZE
UDI-Device Identifier00357599000981
UDI-Public00357599000981
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P160030
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 08/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number71938-01
Device Catalogue Number71936
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/19/2023
Supplement Date Received by Manufacturer08/10/2023
Initial Report FDA Received Date07/10/2023
Supplement Report FDA Received Date08/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/04/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
QUALITY RECORD QR869779
Patient Sequence Number1
Patient SexUnknown
-
-