| Model Number |
FIBERTAG TIGHTROPE WITH FLIPCUTTER III |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
06/07/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event has not yet been determined as the device has not been returned for evaluation at this time.A follow-up report will be submitted, including a most likely cause if a root cause can not be determined.
|
| |
|
Event or Problem Description
|
|
On 06/08/2023, it was reported by a facility representative via phone that an ar-1288rtt-fc3 tightropes flipcutter blade broke off.This was discovered during a case, broken portion of blade unable to be removed, additional surgery required to remove piece.Revision occurred (b)(6) 2023 and was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Complaint allegation was not confirmed.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause(s) for the reported failure can be a user error due to improper bone preparation, misaligned insertion, and/or prying/levering the device during insertion.
|
| |
|