• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number CXA360005
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 06/12/2023
Type of Reportable Event Serious Injury
Event or Problem Description
The following was reported to gore: on (b)(6) 2023, the patient underwent an endovascular procedure in zone 9-infrarenal to treat an aortic aneurysm utilizing gore® excluder® conformable aaa endoprosthesis (aortic extender, trunk ipsilateral leg), and gore® excluder® aaa endoprosthesis (contralateral leg).Reportedly, the aortic extender had to be opened up in the case as the main body's (trunk ipsilateral leg) constraining loop malfunctioned and the main body trunk device had to be pulled down a little.The reason being that one of the stents on the proximal top side of the trunk was flipped or inverted, and not upright.Physician tried pulling it down a bit to get the stent to unflip and straighten but it would not open up as the constraining mechanism broke at which point they put a mob balloon in and inflated it that led to the trunk ipsi being opened back up, however, the device was bit below the renals from the intended landing zone as they had pulled it down and an aortic extender was put in to extend proximally to renals.However, the aortic extender while it landed perfectly at the intended landing zone below the renals, it slid down 15mm into the main body of trunk ipsi leg and the medical team did not want to put in another aortic extender.All devices were implanted successfully, there were no endoleaks.The patient tolerated the procedure.
 
Additional Manufacturer Narrative
According to the gore® excluder® conformable aaa endoprosthesis instructions for use (ifu), potential adverse events that may occur and/or require intervention include, but are not limited to, endoprosthesis or delivery system: component migration.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
H6: code c19 - a review of the manufacturing records indicated the lot met all pre-release specifications.H6: code b20 -the device remains implanted and was therefore not available for engineering evaluation by gore.According to the gore® excluder® conformable aaa endoprosthesis instructions for use (ifu), potential adverse events that may occur and/or require intervention include, but are not limited to, endoprosthesis or delivery system: : component migration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
jaskaran parhar
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key17286911
Report Number3007284313-2023-02609
Device Sequence Number3011247
Product Code MIH
UDI-Device Identifier00733132650972
UDI-Public00733132650972
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P200030
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/20/2025
Device Model NumberCXA360005
Device Catalogue NumberCXA360005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/12/2023
Supplement Date Received by Manufacturer07/11/2023
Initial Report FDA Received Date07/10/2023
Supplement Report FDA Received Date07/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age57 YR
Patient SexFemale
-
-