| Model Number |
5.5MM BC CORKSCREW FT W/ SUTURETAPE |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event has not yet been determined as the device has not been returned for evaluation at this time.A follow-up report will be submitted, including a most likely cause if a root cause can not be determined.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2023, it was reported by a sales representative via email that an ar-1927bct biocomposite corkscrew anchor broke mid-body during insertion.All the broken fragments were retrieved.This was discovered during an rcr procedure on (b)(6), 2023.Additional information received on 6/15/2023: the case was completed using another ar-1927bct biocomposite corkscrew.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely causes of the reported failure are misaligned insertion, and prying/leveraging the device during insertion.
|
| |
|