| Model Number |
3300 |
| Medical Device Problem Codes |
Defective Device (2588); Key or Button Unresponsive/not Working (4063)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this programmer was unresponsive to touch as the touch screen freezes.Additional information received which indicated that the issue did not occur during implant procedure.This programmer remains in service.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for product analysis.Although the device was not returned for analysis, boston scientific performed an investigation with all available information.An unresponsive touchscreen is a known and anticipated potential hazard that can occur during use of the device which is accounted for in device risk documentation and labeling.Boston scientific has received similar reported events where the latitude programmer screen became unresponsive to touch and a complaint review confirmed that the occurrence of harm associated with these events is low.Investigation of this behavior has not found any commonalities (i.E., device serial numbers, date of the event, date of manufacture, etc.) in the reports received.
|
| |
|
Event or Problem Description
|
|
It was reported that this programmer was unresponsive to touch as the touch screen freezes.Additional information received which indicated that the issue did not occur during implant procedure.This programmer remains in service.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for product analysis.Although the device was not returned for analysis, boston scientific performed an investigation with all available information.An unresponsive touchscreen is a known and anticipated potential hazard that can occur during use of the device which is accounted for in device risk documentation and labeling.Boston scientific has received similar reported events where the latitude programmer screen became unresponsive to touch and a complaint review confirmed that the occurrence of harm associated with these events is low.Investigation of this behavior has not found any commonalities (i.E., device serial numbers, date of the event, date of manufacture, etc.) in the reports received.This programmer was thoroughly inspected and analyzed upon receipt at our quality assurance laboratory.The programmer was powered on and it was confirmed that the touchscreen was intermittently unresponsive due to not detecting touch inputs.No error or fault codes were recorded in the programmer's logfile.An unresponsive touchscreen is a known and anticipated potential hazard that can occur during use of the device which is accounted for in device risk documentation and labeling.Boston scientific has received similar reported events where the latitude programmer screen became unresponsive to touch and a complaint review confirmed that the occurrence of harm associated with these events is low.Investigation of this behavior has not found any commonalities (i.E., device serial numbers, date of the event, date of manufacture, etc.) in the reports received.
|
| |
|