• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC EXPEDIUM SFX CROSS CONNECTOR SYSTEM CONNECTOR A3 5.5 X 32-37MM; PEDICLE SCREW SPINAL SYSTEM, ADOLESCENT IDIOPATHIC SCOLIOSIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC EXPEDIUM SFX CROSS CONNECTOR SYSTEM CONNECTOR A3 5.5 X 32-37MM; PEDICLE SCREW SPINAL SYSTEM, ADOLESCENT IDIOPATHIC SCOLIOSIS Back to Search Results
Model Number 189401303
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/15/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthese is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d2: additional procodes: kwp, nkb, mni, kwq, mnh.D9: complainant part is not expected to be returned for manufacturer review/investigation.E3: initial reporter is a synthese employee.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from japan reports an event as follows: it was reported that on (b)(6) 2023, the patient underwent a spinal fusion (t8-l2) for a vertebral body fracture.After temporary fixation of the sfx cross connector was completed, the surgeon attempted final tightening with a sfx torque driver shaft.However, the shaft got stripped multiple times while engaged with the set screw of the sfx cross connector.The final tightening could not be completed, so the surgeon switched from the sfx cross connector to another standard sfx f9.The procedure was completed successfully with no surgical delay.No further information is available.This report is for an expedium sfx cross connector system connector a3 5.5 x 32-37mm.This is report 1 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A review of the receiving inspection (ri) for t-pal proti, 10x9x28 mm, was conducted identifying that lot number om11902 was released in one batch.Batch1: lot qty of (b)(4) units were released on 09 aug 2022 with no discrepancies.Supplier: (b)(4).As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPEDIUM SFX CROSS CONNECTOR SYSTEM CONNECTOR A3 5.5 X 32-37MM
Common Device Name
PEDICLE SCREW SPINAL SYSTEM, ADOLESCENT IDIOPATHIC SCOLIOSIS
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 
SZ  
3035526892
MDR Report Key17289907
Report Number1526439-2023-01246
Device Sequence Number6347698
Product Code OSH
UDI-Device Identifier10705034168036
UDI-Public(01)10705034168036
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K062196
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number189401303
Device Catalogue Number189401303
Device Lot NumberOM11902
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/15/2023
Supplement Date Received by Manufacturer07/11/2023
Initial Report FDA Received Date07/10/2023
Supplement Report FDA Received Date07/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - INSERTION INSTRUMENTS: EXPEDIUM SFX; UNK - LOCKING/SET SCREWS: EXPEDIUM SFX
Patient SexUnknown
-
-