| Model Number |
IPN923488 |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that incoming product inspection, a greasy substance was noted on the mask.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The sample was returned and sent to the manufacturing site for investigation.The manufacturing site reports a "black spot observed on product".A device history record review was performed and no relevant findings were identified.The manufacturer conclusion is as follows: "the root cause of the complaint is "undetermined/unknown" as the origin of the debris claimed unable to be determine whether it was from manufacturing site or not as the product was already removed from the pouch.Therefore, the customer complaint report concludes that there is no further escalation or additional action are required.".
|
| |
|
Event or Problem Description
|
|
It was reported that incoming product inspection, a greasy substance was noted on the mask.No patient involvement.
|
| |
|