| Model Number |
X834H |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 device not returned.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Was there any adverse consequence associated with the patient? did the needle fall into the patient? if yes: was the needle piece(s) retrieved during the same procedure? were x-rays taken to locate the needle piece(s)? what measures were taken to retrieve the broken piece? was there any additional tissue damage as a result of searching for the needle piece? does a piece of the needle remain in the patient¿s tissue? if yes: is there any plan in place to remove the needle piece in the future? if yes, please provide the scheduled date.What tissue structure the broken needle was located? what is the surgeon's opinion of consequences to the patient? a manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a cardiovascular procedure on an unknown date and suture was used.After removal of the pericardial drain, the doctor wanted to perform a stitch using the device.When making the stitch, the needle broke into 2 separate parts.One remained attached to the thread and the other remained in the patient, it was recovered with a kocher forceps.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 9/15/2023.Additional information: h6 this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 9/11/2023.H3 investigational summary: the product received for analysis was identified as product code x834h.Visual inspection and metallurgical analysis were performed on the returned product.A fracture was observed at the suture attachment of the needle.The needle was received in two pieces, only one of the mating fracture surfaces was examined for this evaluation.The needle was noted with marks that appears to be by surgical instrument.A scanning electron microscope was used to examine the fracture surface and surrounding area of the needle.The fracture surfaces were examined in multiple locations to determine the fracture mode.The evaluation of the sample revealed the fracture was composed of microvoid coalescence, which is evidence of a ductile fracture mode.This was a ductile fracture.The evidence of this examination indicates that the breakage occurred at the attachment area of the needle during use due to tensile overload.There is no evidence of any material flaw or defect that would cause premature failure.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Additional information was requested, the following was obtained: was there any adverse consequence associated with the patient? no except for a slightly longer procedure than expected.Were x-rays taken to locate the needle piece(s)? no.Was there any additional tissue damage as a result of searching for the needle piece? no.What tissue structure the broken needle was located? epidermis, the broken end was recovered with kocher forceps.Does a piece of the needle remain in the patient¿s tissue? no.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|