| Model Number |
03.404.022S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/16/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Device report from colombia reports an event as follows: it was reported that during a procedure on (b)(6) 2023, while reaming the medullary channel the tips of the reamer split.The specialist checked with the image intensifier and confirmed that all fragments were removed.No further information is available.This report is for a 13.0mm reamer head for ria 2 sterile.This is report 1 of 2 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d2: additional procode: hrx.D9: complainant part is not expected to be returned for manufacturer review/investigation.E3: initial reporter is a synthes employee.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot: manufacturing location: supplier ¿ (b)(4) / inspected and packaged by: monument; release to warehouse date: 20-oct-2022; expiration date: 01-oct-2032; part number: 03.404.022s, 13.0mm reamer head for ria 2-sterile; lot number: 1491p61 (sterile); lot quantity: (b)(4).Production order traveler met all inspection acceptance criteria.Packaging label log (pll), lmd rev aa was reviewed and determined to be conforming.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.Scn 23706 supplied by sterigenics was reviewed and determined to be conforming.This lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the inspection or release of the product that would contribute to this complaint condition.Component part(s) reviewed: part number: 03.404m022, ria ii reamer head 13.0mm; lot number: 635p113; lot quantity: (b)(4).Inspection instruction met all inspection acceptance criteria.Inspection sheet, incoming final inspection, ns073690 rev b met all inspection acceptance criteria.Certificate of compliance received from mark two engineering dated 08-apr-2022 was reviewed and determined to be conforming.Certification for heat treat supplied to mark two engineering from vacu-braze inc.Dated 31-mar-2022 was reviewed and determined to be conforming.Note: part number notated on certification from vacu-braze for lot number 1007684-003 is incorrect; the part description, however, is correct.Heat treat specified at 50-55 hrc.Results certified at 52-53 hrc.Certificate of analysis supplied to mark two engineering from banner medical dated 28-oct-2020 was reviewed and determined to be conforming.Material certification supplied to banner medical from universal dated 27-jul-2020 was reviewed and determined to be conforming.Device history review: 21-jun-2023: dhr reviewed by: jbucci this lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|