| Model Number |
434900 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Cardiac Arrest (1762)
|
| Date of Event |
05/31/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The pk papyrus covered stent system was selected for treatment of a type iii perforation in the proximal rca.The stent came off balloon inside patient and was not deployed.The stent was pinned by another stent.The procedure was completed successfully.5 days after procedure the patient became hemodynamic instable, needed resuscitation, and died in hospital.Further clarifications are ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
11-aug-2023 the treating physician rated the occurrence as both no device- and no procedure-related complication.The affected pk papyrus stent system was not returned to biotronik and could therefore not be subjected to a technical investigation.Images of the case such as angiographies could also not be obtained.The provided clinical information (pk papyrus device registration form) and the product release documentation were reviewed to establish whether a device deficiency contributed to this event.Review of the product release documentation confirmed that the device was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.Based on the provided documentation and the conducted review no manufacturing related root cause could be identified.The treating physician rated the outcome as both no device- and no procedure-related complication.Pk papyrus is indicated for the treatment of acute perforations of native coronary arteries and coronary bypass grafts in vessels 2.5 to 5.0 mm in diameter.As such, this device is typically used in emergent response to a life-threatening adverse procedural event unrelated to use of the pk papyrus.
|
| |
|
Search Alerts/Recalls
|