It was reported that during the surgery, after fired, noted the staples malformed.Changed another one to continue the surgery , could not fire.Changed the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
|
|
(b)(4).Date sent: 7/11/2023.D4: batch # 533a80.Investigation summary: the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned reload.Visual analysis of the returned sample revealed that the ecr45d reload was returned unfired with the reload pan partially dislodged from the left proximal side.In addition, 12 double drivers and 4 single drivers out of position, making the reload non-functional.No functional test was performed due the condition of the reload.As part of our quality process, the manufacturing records of this batch number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached on what may have caused the reload pan to dislodge. a manufacturing record evaluation was performed for the finished device batch number 533a80, and no non-conformances were identified.
|