| Model Number |
UP501 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/26/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the up501 did not work during the procedure.The procedure had to be aborted and was continued after two days.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information is provided in section d9 to reflect that the product was returned for evaluation.The item in question was returned to the manufacturer.The investigation is not completed yet.The investigation results are pending.The event is filed under internal (b)(6) complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
An endurance test was carried out over 8 hours during which the device was disconnected from the power supply every 2 minutes and then supplied with power again in order to cause the automatic opening and closing of the valve.The opening and closing of the valve was also tested for approximately one hour using the push button on the device and the foot switch.No irregularities could be detected.The described error could not be reproduced.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|