| Model Number |
24700 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that catheter entrapment occurred.The patient underwent angioplasty of the 50% stenosed target lesion with mixed morphology located in the mildly to moderately tortuous and mildly calcified iliac artery.After gaining vascular access via the right common femoral artery with a 6fr sheath, a 300cm, 8cm v-18 control wire guidewire was advanced deep into the aorta and into the target lesion.A 7.0mmx40mmx135cm (4f) sterling balloon catheter was prepped and advanced over guidewire into the external iliac for pre-dilation.The balloon was inflated to 10 atmospheres and worked perfectly; however, during withdrawal, the balloon was stuck on the guidewire.Flushing, inflating, and deflating the balloon was tried but it would not come off the guidewire.The balloon and guidewire were pulled out intact together through the sheath and a new guidewire was used to access the lesion.The procedure was completed successfully.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that catheter entrapment occurred.The patient underwent angioplasty of the 50% stenosed target lesion with mixed morphology located in the mildly to moderately tortuous and mildly calcified iliac artery.After gaining vascular access via the right common femoral artery with a 6fr sheath, a 300cm, 8cm v-18 control wire guidewire was advanced deep into the aorta and into the target lesion.A 7.0mmx40mmx135cm (4f) sterling balloon catheter was prepped and advanced over guidewire into the external iliac for pre-dilation.The balloon was inflated to 10 atmospheres and worked perfectly; however, during withdrawal, the balloon was stuck on the guidewire.Flushing, inflating, and deflating the balloon was tried but it would not come off the guidewire.The balloon and guidewire were pulled out intact together through the sheath and a new guidewire was used to access the lesion.The procedure was completed successfully.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of a sterling balloon catheter with a guidewire stuck inside the device.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed buckling to the guidewire lumen 4cm and 8cm from the tip.Microscopic examination revealed no additional damages.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Search Alerts/Recalls
|