| Model Number |
866389 |
| Medical Device Problem Code |
Application Network Problem (2879)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A follow-up report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
A customer biomedical engineer (biomed) reported that all of patient monitoring keeps rebooting and disconnecting from server in all floors and departments.The biomed called back and spoke to a philips remote service engineer (rse); some of servers and switches for multiple departments are still not working or is intermittently.An additional follow-up call indicated the site was running according to the biomed.The cause of the issue is unknown and hospital information technology (it) is also working with philips on the issue.Further information will be reviewed to determine the root cause.The device was in clinical use at the time the issue was discovered.There was no adverse event or patient harm reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The site had a philips field service engineer (fse) visit on 23jun2023.It was found that the server became unreachable on the customer supplied clinical network (cscn).Based on the information available and the testing conducted, the cause of the reported problem was due to the customer network disconnecting from the server.Cscn issues are not considered a reportable event, as this is not considered a philips product.The investigation concludes that no further action is required at this time.
|
| |
|