MAUDE Adverse Event Report: AVANOS MEDICAL INC. HOMEPUMP C-SERIES, 270 ML, 5 ML/HR(DEHP); ELASTOMERIC LFR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. HOMEPUMP C-SERIES, 270 ML, 5 ML/HR(DEHP); ELASTOMERIC LFR Back to Search Results
Model Number C270050
Medical Device Problem Code Excess Flow or Over-Infusion (1311)
Health Effect - Clinical Code Hot Flashes/Flushes (2153)
Date of Event 07/14/2023
Type of Reportable Event Malfunction
Event or Problem Description
Avanos medical received a single report that referenced seven different incidences, which were associated with separate units, involving seven different events.This is the first of seven reports.Refer to 2026095-2023-00071 for the second event.Refer to 2026095-2023-00072 for the third event.Refer to 2026095-2023-00073 for the fourth event.Refer to 2026095-2023-00074 for the fifth event.Refer to 2026095-2023-00075 for the sixth event.Refer to 2026095-2023-00076 for the seventh event.Fill volume: unknown, flow rate: unknown, procedure: unknown, cathplace: unknown, infustion start time: (b)(6) 2023 11:07am, infection stop time: (b)(6) 2023 08:30am.It was reported the pump infused faster than the 46-hour timepoint, and when the pump runs fast the patient experiences flushing as a side effect.There was no reported injury or medical intervention.
 
Additional Manufacturer Narrative
The actual device is reported to be available but has not been returned for evaluation.A review of the device history record is in-progress.All information reasonably known as of 11-jul-2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
The device history record for the reported lot number, 30234039, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements.All information reasonably known as of 22-sep-2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
The device was evaluated.The evaluation summary noted the sample did not show any infusion on the glass orifice portion of the pump while verifying the infusion in different locations of the tubing.After rehabilitating the tubing by cutting multiple areas to identify the location of occlusion/ blockage, the area of occlusion/ clogging was in the flow restrictor (glass orifice).An unknown substance was unable to dissolve.The filter was also exhibited to leak.Reported "fast flow" was not confirmed.Root cause could not be determined.All information reasonably known as of 22-nov-2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
The actual complaint product was returned for evaluation.All information reasonably known as of 11-oct-2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMEPUMP C-SERIES, 270 ML, 5 ML/HR(DEHP)
Common Device Name
ELASTOMERIC LFR
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S. DE R.L. DE C.V.
ave noruega edificio d-1b
fraccionamiento rubio
tijuana b.c.
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key17302955
Report Number2026095-2023-00070
Device Sequence Number20504346
Product Code MEB
UDI-Device Identifier00193494135683
UDI-Public00193494135683
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K052117
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 11/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/24/2025
Device Model NumberC270050
Device Catalogue NumberN/A
Device Lot Number30234039
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/16/2023
Supplement Date Received by Manufacturer09/01/2023
09/18/2023
12/06/2023
Initial Report FDA Received Date07/11/2023
Supplement Report FDA Received Date09/22/2023
10/12/2023
11/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/21/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5FU.
Patient SexMale
-
-