| Model Number |
67SP045 |
| Medical Device Problem Code |
Stretched (1601)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to use, the cuff was tested and found to be not symmetrical.Customer states this could cause pressure to the trachea.The device was not used on the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was received without the original packaging.Visual inspection noted no unusual conditions.Functional testing observed that the balloon inflated as expected fully and symmetry value was 40 percent greater than 33 percent as outlined in manufacturing specifications.The complaint was not confirmed, and no other analysis was performed.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.This issue will be monitored, and further actions taken accordingly.
|
| |
|