| Model Number |
FINETUNER ECHO [FINETUNER 2] |
| Medical Device Problem Codes |
Circuit Failure (1089); No Device Output (1435)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
|
| |
|
Event or Problem Description
|
|
A finetuner echo was returned to service and repair because of being defective.No injury has been reported.Most likely the device was not handed out to any user.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: according to the information received, a finetuner echo was sent to service _ repair due to it not functioning.During device investigation a failed capacitor was detected which was clearly the reason for the reported issue.Electrical inspection could not be performed because of the observed malfunction.No injury was reported.This is a final report.
|
| |
|
Event or Problem Description
|
|
A finetuner echo was returned to service and repair because of being defective.No injury has been reported.
|
| |
|
Search Alerts/Recalls
|