MEDOS INTERNATIONAL SàRL CH EXPEDIUM VERSE SPINE SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 7.0 X 45MM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Model Number 199723745S
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/19/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional device product codes: kwq, nkb.Complainant part is not expected to be returned for manufacturer.Review/investigation.Reporter is a j&j employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in japan as follows: it was reported that during a planned second spinal surgery (t12-l3) on (b)(6) 2023, when attempting to connect t12-l3, the locks of the heads of the verse fenestrated screws were not released, and the connection using rods could not be applied.The fenestrated screws had been implanted in the first surgery on (b)(6) 2023.There were four places where the locks of the heads of the verse fenestrated screws had to be released, therefore the four screws were replaced with sized-up ones, and the connection using rods was done.The surgery was completed successfully with no delay.This report involves one expedium verse spine system fenestrated cortical fix polyaxial screw 7.0 x 45mm.This is report 1 of 4 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPEDIUM VERSE SPINE SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 7.0 X 45MM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
SZ  
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 
SZ  
3035526892
MDR Report Key17304927
Report Number1526439-2023-01258
Device Sequence Number12358368
Product Code KWP
UDI-Device Identifier10705034444505
UDI-Public(01)10705034444505
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K173095
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number199723745S
Device Catalogue Number199723745S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/19/2023
Initial Report FDA Received Date07/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5.5 EXP VERSE FEN SCR 7.0X45; 5.5 EXP VERSE FEN SCR 7.0X45; 5.5 EXP VERSE FEN SCR 7.0X45; UNKNOWN RODS
Patient SexUnknown
-
-