MAUDE Adverse Event Report: ARGON MEDICAL DEVICES L-CATH PICC D/L 18GA (3.5F) 1.20MM X 60CM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARGON MEDICAL DEVICES L-CATH PICC D/L 18GA (3.5F) 1.20MM X 60CM Back to Search Results
Model Number 384467
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/19/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
According to the product experience report, there was no sample to be returned.Additionally, no photographic or visual evidence of any kind was provided which would have allowed for the allegation to be reviewed.Without such evidence, this complaint could not be confirmed and determining a definite root cause and corrective action is not possible.If the sample is returned at a future date, a follow-up report will be submitted.
 
Event or Problem Description
Called to bedside by rn who reported fluids leaking from picc.Upon assessment, lipids leaking at microclave connection.Connection appeared loose, upon tightening, blood began to back up into lumen.Of note, cap change performed previous day on this red lumen and t-connector applied due to history of this type of picc lines cracking, however, t-connector had been applied with microclave attached to picc line.Microclave removed under sterile conditions and new t-connector applied.Crack confirmed, line clamped and fluids d/c'd from line.Flc notified, us guided piv placed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
L-CATH PICC D/L 18GA (3.5F) 1.20MM X 60CM
Common Device Name
L-CATH PICC
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek rd
athens TX 75751
Manufacturer Contact
elnaz rahman
1445 flat creek rd
athens, TX 75751
9036759321
MDR Report Key17306709
Report Number0001625425-2023-01120
Device Sequence Number20429051
Product Code LJS
UDI-Device Identifier00886333209866
UDI-Public00886333209866
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K091670
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/28/2025
Device Model Number384467
Device Catalogue Number384467
Device Lot Number11418339
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/29/2023
Initial Report FDA Received Date07/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/27/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-