|
According to the product experience report, there was no sample to be returned.Additionally, no photographic or visual evidence of any kind was provided which would have allowed for the allegation to be reviewed.Without such evidence, this complaint could not be confirmed and determining a definite root cause and corrective action is not possible.If the sample is returned at a future date, a follow-up report will be submitted.
|
|
Called to bedside by rn who reported fluids leaking from picc.Upon assessment, lipids leaking at microclave connection.Connection appeared loose, upon tightening, blood began to back up into lumen.Of note, cap change performed previous day on this red lumen and t-connector applied due to history of this type of picc lines cracking, however, t-connector had been applied with microclave attached to picc line.Microclave removed under sterile conditions and new t-connector applied.Crack confirmed, line clamped and fluids d/c'd from line.Flc notified, us guided piv placed.
|