MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE PLUS TMZF HIP STEM #4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE PLUS TMZF HIP STEM #4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 6021-0435
Medical Device Problem Codes Degraded (1153); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Joint Laxity (4526); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/17/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.H3 other text: not returned to the manufacturer.
 
Event or Problem Description
On (b)(6) 2008, implantation of a stryker hip prosthesis with trident screw acetabular cup, pe inlay, accolade stem, metal head.In the course now increasing instability with multiple falls.On (b)(6) 2023, evidence of premature material wear.V.A implant fracture, revision surgery on (b)(6) 2023 with massive metallosis and evidence of complete wear of the metal femoral neck.
 
Event or Problem Description
On (b)(6) 2008 implantation of a stryker hip prosthesis with trident screw acetabular cup, pe inlay, accolade stem, metal head.In the course now increasing instability with multiple falls.On 5/17/2023 evidence of premature material wear.V.A implant fracture, revision surgery on (b)(6) 2023 with massive metallosis and evidence of complete wear of the metal femoral neck.
 
Additional Manufacturer Narrative
An event regarding wear and disassociation involving an accolade stem that was mated with a lfit v40 cocr head was reported.The event was confirmed.Method & results: device evaluation and results: the reported device was not returned however photographs were provided for review.The photographs show a recently explanted head, stem, shell and liner with body fluids on it.The head seemed to be removed still seated onto the trunnion of the stem.However, once removed from the stem in a second photo, there is evidence of wear of the trunnion, black tissue and the taper of the head appears dark in colour.Clinician review: a review of the provided medical records by a clinical consultant indicated: 'this case concerns a patient who underwent a primary cementless total hip arthroplasty in 2008 and then approximately 18 years later developed head neck disassociation with metallosis requiring revision surgery.I can confirm that the patient underwent a primary total hip replacement and sustained a head neck disassociation since i was able to review x-rays showing the implants.I cannot confirm the revision surgery since i have no documentation such as office notes, operation note or post revision x-rays.The root causes of had neck disassociation with metallosis are multifactorial including surgical technique factors, trunnion and head preparation prior to implantation, patient factors such as activity level and bmi, and implant factors.The explanted femoral prosthesis and femoral head should be submitted to stryker engineers for examination and evaluation.' device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: no other similar events were reported for the lot.Conclusion: the reported accolade stem was mated with a lfit v40 cocr head.The cocr head has been identified to be within scope of nc and capa.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE PLUS TMZF HIP STEM #4
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key17307041
Report Number0002249697-2023-00768
Device Sequence Number15269281
Product Code LPH
UDI-Device Identifier04546540510693
UDI-Public04546540510693
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2013
Device Catalogue Number6021-0435
Device Lot Number26065502
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/16/2023
Supplement Date Received by Manufacturer08/18/2023
Initial Report FDA Received Date07/12/2023
Supplement Report FDA Received Date09/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/19/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age86 YR
Patient SexFemale
Patient Weight65 KG
-
-