MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RETROGRADE DRL 8MM; BIT, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RETROGRADE DRL 8MM; BIT, SURGICAL Back to Search Results
Model Number 72204042
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/15/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unknown procedure, the doctor went to use the retrograde drill, before he activated the drill to make the tunnel of the femur, it broke and the piece was removed from the patient.The procedure was successfully completed with no surgical delay using a back-up device.No further complications were reported.
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H10: internal complaint reference (b)(4).H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.Insufficient product identification information was provided and thus a manufacturing record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.H11: corrected information in d8.
 
Additional Manufacturer Narrative
The reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.The device was returned with the entire antegrade drill with the retrograde drill inside is fractured off the shaft.The actuator wire is fractured with the end protruding from the shaft twisted and bent.The proximal end of the actuator wire is jammed under the bar on the grey slider.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with unintended use of the device.Factors that could have contributed to the failure include an application of excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RETROGRADE DRL 8MM
Common Device Name
BIT, SURGICAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key17307849
Report Number1219602-2023-01136
Device Sequence Number7158162
Product Code GFG
UDI-Device Identifier00885554037715
UDI-Public00885554037715
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72204042
Device Catalogue Number72204042
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/19/2023
Supplement Date Received by Manufacturer11/15/2023
02/21/2024
Initial Report FDA Received Date07/12/2023
Supplement Report FDA Received Date11/16/2023
02/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age37 YR
Patient SexMale
-
-