| Model Number |
00884450030776 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that the stent fractured approximately 45 days after implantation.The fracture required the removal of the stent.The patient is doing well; however, the disease is no longer being treated.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was returned for evaluation.The complaint was confirmed.No stent fracture was seen as nitinol metal frame was intact, but holes were seen in the stent.Holes seemed to be present in areas with excessive tissue growth.Note: tissue growth is dependent on each individual patient and the stage/condition of their malignancy.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
|
| |
|