| Model Number |
IPN030312 |
| Medical Device Problem Code |
Deformation Due to Compressive Stress (2889)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the guidewire was bent inside the packaging.No patient involvement with device was reported.A new kit was obtained for use.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The actual device was not returned; however, the customer provided one photo for analysis.The complaint of a kinked guide wire was able to be confirmed by the photo.A complete visual inspection could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.The instructions for use (ifu) provided with this kit warns the user, "do not use if package is damaged." the customer report of a kinked guide wire was confirmed by visual inspection of the customer supplied photo.However, full complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported the guidewire was bent inside the packaging.No patient involvement with device was reported.A new kit was obtained for use.
|
| |
|