| Model Number |
15 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/16/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and verified the reported issue.Stryker determined that the cause of the reported issue was due to a failed system pcb assembly.A replacement for this part is no longer available due to part obsolescence.The device was returned to the customer unrepaired.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report a non-critical issue with their device.There was no report of patient use associated with the reported event.Upon evaluation of the customer's device, stryker observed that the device is not completing the boot up process.As a result, defibrillation therapy may be unavailable, if needed.
|
| |
|