| Model Number |
IPN917285 |
| Medical Device Problem Code |
Power Problem (3010)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/16/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The report states, "pump had purge failure alarm and powered down twice while on patient".At the time of this report, the customer has not returned our multiple requests for additional information.If further information is received, the complaint file will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).Other remarks: n/a.Corrected data: n/a.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Returned for investigation were the martek power supply (p/n: 33-0063-001, s/n: (b)(6) and the battery (p/n: 4000-9022-001, l/n: 18c21c0245).The samples were returned in a brown cardboard shipping box with protective packaging.Visual inspection of the power supply and battery was performed and no abnormality was noted.The power supply and battery were installed into a known good ac3 for functional testing.The system powered-up successfully.The power supply voltage check was performed per ac3 series service manual and all output voltages were within specifications.The battery load test was performed.The iabp was run on battery until the iabp shut down.The battery was then left to charge in the iabp for over 9 hours.The full charge of the battery was 13.3 volts.The iabp gave a proper alarm when disconnected from the ac power.Pumping was initiated.The unit ran for over 90 minutes along with the proper alarms "less than 20, 10, and 5 minutes remaining".A device history record (dhr) review was conducted for the lot number/serial number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint of "powered down twice" is not confirmed.The returned power supply and battery passed visual and functional test specifications.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.
|
| |
|
Event or Problem Description
|
|
The report states, "pump had purge failure alarm and powered down twice while on patient".At the time of this report, the customer has not returned our multiple requests for additional information.If further information is received, the complaint file will be updated.
|
| |
|