ROCHE DIAGNOSTICS LIPC; LIPASE-ESTERASE, ENZYMATIC, PHOTOMETRIC, LIPASE Back to Search Results
Model Number LIPC
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2023
Type of Reportable Event Malfunction
Event or Problem Description
The initial reporter stated they received discrepant results for one patient sample tested with the lipc (lipase colorimetric assay) assay on a cobas 4000 c311 stand alone system (serial number (b)(6).The initial reporter also mentioned random quality control errors.The sample initially resulted in an lipc value of 83 u/l, accompanied by a data flag.The sample was repeated resulting in a value of 30 u/l.No questionable result was reported outside of the laboratory.
 
Additional Manufacturer Narrative
The field service engineer ran instrument performance testing which was out of specifications.A misadjusted gear pump and sample probe were found.After the gear pump, the sample probe position and the rinse arm flow were adjusted, instrument performance testing was repeated, and was within specifications.The investigation determined the service actions performed resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIPC
Common Device Name
LIPASE-ESTERASE, ENZYMATIC, PHOTOMETRIC, LIPASE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key17309419
Report Number1823260-2023-02231
Device Sequence Number6384418
Product Code CHI
UDI-Device Identifier04015630917570
UDI-Public04015630917570
Combination Product (Y/N)N
Initial Reporter CountryEG
PMA/510(K) Number
K972250
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2023
Device Model NumberLIPC
Device Catalogue Number03029590322
Device Lot Number66460101
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2023
Supplement Date Received by Manufacturer08/08/2023
Initial Report FDA Received Date07/12/2023
Supplement Report FDA Received Date08/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-