SYNTHES GMBH CHUCK WITH KEY-REAM SPEED F/BPL; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
530.732 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the chuck with key device broke in half during use.It was not reported if the event occurred during a surgical procedure.It was not reported if there was a delay to a planned procedure.It was not reported if there was a spare device available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the chuck with key device coupling was fractured.It was further determined that the device failed pre-tests for general condition, check drill chuck function, check cannulation of attachment, and check the handpiece coupling.Therefore, the reported condition that the device broke in half during use was confirmed.The assignable root cause was determined to be traced to the user, which is user error.
|
| |
|