| Model Number |
0210 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
Unspecified Respiratory Problem (4464)
|
| Date of Event |
06/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Tubing becomes disconnected from the ventilator.
|
| |
|
Additional Manufacturer Narrative
|
|
The nebulizer's failure to release aerosol and the tubing popping off would delay treatment to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The nebulizer's failure to release aerosol and the tubing popping off would delay treatment to the patient.Complaint is not confirmed.Testing of a different lot from inventory was found and units failed so ncmr-03535 was initiated.Reviewed the dhr and no quality issues were identified.Performed a risk analysis with rma-20013f and determined a severity rating of 6/10.Sent a resolution to the customer.
|
| |
|
Event or Problem Description
|
|
Tubing becomes disconnected from the ventilator.
|
| |
|