• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO MANUFACTURING USA, LLC STAR S4 IR; EXCIMER LASER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO MANUFACTURING USA, LLC STAR S4 IR; EXCIMER LASER SYSTEM Back to Search Results
Model Number 0030-4864
Medical Device Problem Code Unintended System Motion (1430)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation: the investigation by the field service engineer resulted in no problems found.The system meets specifications and no product deficiency is present.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
 
Event or Problem Description
Customer reported that their star laser chair moved without any input.The chair was locked and doctor was ready to treat.He pressed the foot pedal after the machine verified iris registration, and it would act like it was going to fire, but it would stop treatment due to losing focus/tracking because the patient was moving down from the desired spot.
 
Additional Manufacturer Narrative
Section h4.Device manufacture date corrected to dec 29, 2006.Additional information: manufacturing record review was performed.There were no discrepancies or non-conformances experienced during manufacturing.The system and its components met all specifications prior to being released.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STAR S4 IR
Common Device Name
EXCIMER LASER SYSTEM
Manufacturer (Section D)
AMO MANUFACTURING USA, LLC
510 cottonwood drive
milpitas CA 95035
Manufacturer Contact
somyata nagpal
1700 e st andrew place
santa ana, CA 92705
7142478552
MDR Report Key17313272
Report Number3012236936-2023-01752
Device Sequence Number6225869
Product Code LZS
UDI-Device Identifier05050474573413
UDI-Public(01)05050474573413
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P910062
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 08/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0030-4864
Device Catalogue Number0030-4864
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/13/2023
Supplement Date Received by Manufacturer07/26/2023
Initial Report FDA Received Date07/12/2023
Supplement Report FDA Received Date08/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/29/2006
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-