|
SYNTHES GMBH DRILL ATTACHMENT MQC FOR PEN DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
05.001.030 |
| Medical Device Problem Codes |
Complete Loss of Power (4015); Excessive Heating (4030)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the netherlands that during service and evaluation, it was determined that the drill attachment device produced heat and did not function.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2023.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: this device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.However, since the investigation is still on-going, the assignable root cause could not be determined at this time.Once investigation has been completed, a supplemental medwatch will be submitted accordingly.Udi: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|