| Model Number |
4K SYNERGYUHD4 BB CAMERA HEAD,AUTOCLAV |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/16/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event has not yet been determined as the device has not been returned for evaluation at this time.A follow-up report will be submitted, including a most likely cause if a root cause can not be determined.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2023, it was reported by a sales representative via sems that (qty.3) of an ar-3210-0029 4k synergyuhd4 bb camera head would go black when pressing the wand in the middle of the case and then it would not give a clear image after.The case was completed successfully by using another camera head.This was discovered during an unspecified procedure, with no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: h6: (head cable) the evaluation determined that the reported event was caused by a defective head cable.(ncr-19324).
|
| |
|